E-023-2-E
Diagnosis results are as follows (phase difference received from the DSP is monitored for 30 seconds,
and the difference between the maximum and minimum values is examined):
・
"Best condition": 25 or less
・
"Good condition": More than 25 and equal to or less than 75.
・
"Not so good condition": More than 75 and equal to or less than 150.
・
"Bad condition": More than 150, or "P.O. Signal Alarm, Drive Input Out of Range" occurred.
⑥
Click the "OK" button to return to the window from before the check, and then click the "Cancel" button
to close the window.
Clicking "OK" here will start the static device installation state check over again.
3.7.1.2.2 Dynamic device installation state check (Dynamic)
①
Click the "Diag/Service (T)" menu, select "Installation" from the "Test/Status" drop-down list, and then
click "Dynamic" under that.
②
The following window will be displayed (Fig. 112). This will check the dynamic device installation state.
Click "OK" after completely stopping the fluid.
③
Click the "OK" button to display a message box (Fig. 113).
④
During the check, a progress bar will appear as shown in Fig. 114, allowing for the confirmation of the
state of the check.
Fig.111
Fig.112
Fig.113
Fig.114