E-023-2-E
2
⑦
Click the "OK" button after all settings are completed to display the message box (Fig. 40).
Click "OK" here to change the settings to the inputted values, and to reflect the changed setting values
in the flowmeter
’
s output. For the sake of safety, if the flowmeter
’
s output is used to control valves or
other such parts, then change that control loop to manual control so that the control loop is in a state
whereby it is not influenced by the flowmeter
’
s output.
⑧
Click "OK" to change the settings to the selected values, and then click the "OK" button in the
displayed message box (Fig. 41) to complete the setting process.
⑨
After the "OK" button is clicked, the item input window will appear again. Click the "Cancel" button to
close the item input window.
To cancel the settings, click the "Cancel" button in any step from
②
to
⑦
.
3.6.2.7 Error output level settings (Error output)
①
Click the "Setup (S)" menu, and then select and click "Error output" from the "Outputs" drop-down list.
②
The following window will be displayed (Fig. 42). Set the analog and pulse output level for when errors
occur here.
Fig.40
Fig.41
Fig.42