E-023-2-E
③
Click the "OK" button to display a message box (Fig. 97).
This conducts checks, so if the output of the flowmeter is being used to control valves and so on, make
sure to switch that control loop to manual control so that it is in a state whereby it is not affected by the
output of the flowmeter.
④
During the check, a progress bar will appear as shown in Fig. 98, allowing for the confirmation of the
state of the check.
⑤
Click "OK" to execute the checks. Once they are finished, a message box (Fig. 99) will be displayed.
Click the "OK" button to complete the hardware check.
The results are displayed as shown in Fig. 100.
⑥
Click the "OK" button to return to the window from before the check, and then click the "Cancel" button
to close the window.
Clicking "OK" here will start the drive resistance check over again.
Fig.97
Fig.98
Fig.99
Fig.100