3
3.6.6 Converter display settings (LCD)
3.6.6.1 Display order settings (Var. priority)
①
Click the "Setup (S)" menu, and then select and click "Var. priority" from the "LCD" drop-down list.
②
The following window will be displayed (Fig. 66). Set the order to be used for displaying measurement
values on the LCD here.
③
If the LCD
’
s "Font" setting is "Double Angle", then two items will be displayed at a time, starting with
the first item. If the setting is "Normal", then three items will be displayed on the screen at a time, so
pick the items you want to see first and assign the numbers accordingly. An item with the setting 0 is
not displayed.
If a number is duplicated or skipped, then an input error will occur and the setting will not be possible.
④
After all settings are complete, click the "OK" button to display a message box (Fig. 67).
⑤
Click "OK" to change the settings to the inputted values, and then click the "OK" button in the displayed
message box (Fig. 68) to complete the setting process.
⑥
After the "OK" button is clicked, the item input window will appear again. Click the "Cancel" button to
close the item input window.
To cancel the settings, click the "Cancel" button in any step from
②
to
④
.
Fig.66
Fig.67
Fig.68