Setting the whole setup to defaults can be accomplished by typing SETUP DEFAULTS. Setting just
measurement one to defaults can be done via M1 SETUP DEFAULTS.
17.5
Measurements
To view the setup and the last reading made measurement one, type M1. Use M2 for measurement two,
etc.
To view all the last measured readings, type LAST. It will show a reading for every active sensor. Typing
LAST does not initiate new measurement.
To initiate a new measurement for each active sensor, type MEAS. The station will measure each sensor
one at a time and display results as it goes along.
If you are interested in only one specific measurement, try M1 LAST and M1 MEAS. Those will show the
last measured value and initiate a new measurement, respectively
To change the current reading of a sensor, type M1 = 12.5. This will change the
Measurement Offset
such
that the said measurement reads the user entered value. For example, if my water level sensor, once set
up, read 3.50, and I knew the water level was at 1.50, I would type M1 = 1.50. The next it was measured,
M1 would read 1.50 (assuming the level does not change). The
Measurement Offset
would have changed
from 0 to -2.00.
17.6
Recording
To start the station, click the Start button in LinkComm, or type RECORDING = ON.
17.7
Downloading the Log
The system will save the measured data in its flash memory each time a measurement is made. The
command LOG command will start a transfer of the log to the connected device.
Use NH to exclude the header; Use NE to omit events.
Help Log gives details on downloading the log.