OMNI 4000/7000 Operations and Maintenance Guide
– Rev F
7-32
Maintenance
16.
a. If you have trouble calibrating the
output channel, reset the channel by
clicking the
Defaults
function key
and attempt to calibrate the
channel again.
b. Press the
Yes
function key to finalize
the calibration of the output channel.
c. Press the
No
function key to discard
the calibration changes.
17.
Continue calibrating additional output
channels by repeating Steps 1 through 16.
7.3.3 Calibration Verification
Input Verification
To verify that the input calibration variables are correct or within the acceptable error tolerance
and so do not require calibration, follow these instructions:
1.
a. From the main menu of the front
panel, select
Utilities
and press
OK
.
b. Select
Calibration
, press
OK
.
c. Select
Input Verification
.
2.
Use the arrow navigation keys to select an
option on the
Input
bar; then press
OK.
The
Process Input Verification
screen includes a horizontal Input bar at the top
with these options:
•
PI allows you to calibrate by process input channel number selection
(for example, 1 through 32).
•
Numbers 1 through 6 allow you to calibrate using a process input variable name
for individual Meters 1 through 6 (for example, Meter 1 Temperature or Meter 5
Density Pressure).
•
S1 and S2 allow you to calibrate using a process input variable name for
Individual Station 1 and Station 2 (for example, Station 1 Density Temperature
or Station 2 Density Pressure).