OMNI 4000/7000 Operations and Maintenance Guide
– Rev F
7-31
Maintenance
10.
Use the front panel to key in the value that
appears on the ammeter:
a. Select the
Measured Output (mA)
?
field.
b. Press
OK
.
c. Enter the value.
d. Press
OK
again or the
Accept
function key.
11.
If needed, press the
Abort
function key to
cancel the calibration.
To continue to the 20 mA calibration
screen, press the
Apply Low
function key.
The next screen shows the
Output Value
(mA)
field default of 20.00.
12.
Wait 20 to 30 seconds for the readings to
stabilize.
13.
Note the value indicated on the ammeter.
14.
Use the front panel of the flow computer to
key in the value displayed by the
ammeter:
a. Select the
Measured Output (mA)
?
field.
b. Press
OK
.
c. Enter the value.
d. Press
OK
again or the
Accept
function key.
15.
If needed, press the
Abort
function key to
cancel the calibration.
To continue to the
Finalize Calibration
screen, press the
Apply HI
function key.