OMNI 4000/7000 Operations and Maintenance Guide
– Rev F
7-29
Maintenance
20.
Press the
Apply HI
function key to
complete the span adjustment or the
Abort
function key to discard.
21.
After pressing the
Apply HI
function key,
the
Finalize Calibration?
screen displays.
a. If you have trouble calibrating the
input channel, reset the channel by
pressing the
Defaults
function key;
then attempt to calibrate the
channel again.
b. Press the
Yes
function key to finalize
the calibration of the input channel.
c. Press the
No
function key to discard
the calibration changes.
22.
Continue calibrating additional input
channels as needed by repeating
these steps.
Output Calibration
To calibrate the 4
‒20 mA output channels, follow these instructions:
1.
a. From the main menu of the front
panel, select
Utilities
and press
OK
.
b. Select
Calibration
, press
OK
.
c. Select
Output Calibration
.
2.
Press
OK
to enter the edit mode of the
Selected Output?
field.