Software Upgrade and Maintenance 6-9
NOTE:
1.
Before uploading the log, make sure the ultrasound system has been connected
to network, otherwise, the system may warn “could not connect to server”
2.
The server (smtp.163.com
) is already specified by the system, the user doesn’t
have to select it.
6.6
Display Parameter Setting
NOTE:
In normal condition, when changing display assembly or main board assembly, the
hardware will update the display data automatically. If synchronization fails, please
load display parameters manually to assure image effect.
1. Log in as Service user.
2. Enter [Maintenance] menu in Setup menu. Then select [Test Main Monitor] to enter.
3. Click t
he upper icon in the screen (prompts “Update” if you put the cursor on it), the update step
is carried out, if fails, please send to R&D center.