Troubleshooting of Regular Malfunctions 10-5
The image abnormality may be caused by
the swivel of the 4D probe or the 4D drive
board. It is suggested to replace the probe
first, and then replace the 4D drive board.
Replace the probe, or
replace the 4D drive board
5
ECG
abnormality
Failed to recognize ECG. It may cause by
the ECG module or the IO board.
Replace the ECG module
or the IO board
ECG wave abnormality
Replace ECG module
10.4 Probe Socket System Malfunction
10.4.1 Module or Board Related
No.
Descriptions
Remarks
1
Main board
10.4.2 Key Points Supporting Troubleshooting
No.
Key Points Supporting Troubleshooting
Remarks
1
Probe recognition of all models of probes when
connecting on the same or different ports
2
Probe board ID of the system
10.4.3 Troubleshooting
No. Fault
Description
Cause Analysis
Measure
1
Probe cannot
be
recognized
This probe cannot be recognized by connecting to all
probe sockets, while recognition of other probes is
normal. Probe malfunction.
Replace the probe.
This probe cannot be recognized when connecting to a
certain socket. Probe board malfunction.
Replace the probe
board
All probes cannot be recognized in all probe sockets;
Probe board ID cannot be read correctly by the system
information;
Main board or probe board malfunction.
Replace the main
board or probe
board.
Содержание Z6
Страница 2: ......
Страница 14: ......
Страница 16: ...2 2 Product Specifications Left side 9 4 7 5 6 8 ...
Страница 72: ......
Страница 119: ...Structure and Assembly Disassembly 7 35 图 7 59 Disassembly of Speaker 8 ...
Страница 120: ......
Страница 132: ......
Страница 142: ......
Страница 152: ......
Страница 168: ......
Страница 171: ...Phantom Usage Illustration B 3 ...
Страница 172: ......
Страница 173: ...P N 046 010452 00 5 0 ...