E-4 Electrical Safety Inspection
LIMITS
For BF
applied parts:
100 μA in Normal Condition.
500 μA in Single Fault Condition.
E.8
Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage using a limiting
resistance, to selected applied part terminals. Current measurements are then taken between the
selected applied part and earth. Measurements are taken with the test voltage (110% of mains) on
applied parts in the normal and reverse polarity conditions.
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity;
Reversed Polarity.
LIMITS
For BF
applied parts: 5000 μA.
E.9
Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining
Applied Part connectors. All measurements may have a true RMS response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
normal polarity (Normal Condition);
reverse polarity (Normal Condition);
normal polarity with open neutral (Single Fault Condition);
reverse polarity with open neutral (Single Fault Condition);
normal polarity with open earth (Single Fault Condition);
reverse polarity with open earth (Single Fault Condition).
LIMITS
For BF
applied parts,
100 μA in Normal Condition.
500 μA in Single Fault Condition.
NOTE:
Make sure the safety analyzer is authorized and complies with the requirements of IEC61010-1.
Follow the analyzer manufacturer's instructions.
Содержание Resona 7
Страница 2: ......
Страница 24: ......
Страница 232: ......
Страница 278: ......
Страница 320: ...12 22 Setup Click I Accept Select I do not want to join the program at this time and click Next...
Страница 326: ......
Страница 386: ......
Страница 396: ......
Страница 424: ......
Страница 442: ......
Страница 451: ...P N 046 007807 02 3 0...