13-30 Probes and Biopsy
Needle-guide Bracket Installation (NGB-024)
1. Put on the sterile transducer sheath.
2. Select a proper needle-guided bracket, and match the groove with the tab of the transducer. Mount
the bracket onto the transducer.
3. Rotate the clamping nuts of the guided bracket on right and left side to fix the bracket and the
transducer. Rotate the needle-type adjusting nut to the ultimate position as shown in the figure.
4. Hold the transducer. Press the biopsy needle to separate needle guided V-shaped block from
pressure position of the needle.
5. Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.
Содержание Resona 7
Страница 2: ......
Страница 24: ......
Страница 232: ......
Страница 278: ......
Страница 320: ...12 22 Setup Click I Accept Select I do not want to join the program at this time and click Next...
Страница 326: ......
Страница 386: ......
Страница 396: ......
Страница 424: ......
Страница 442: ......
Страница 451: ...P N 046 007807 02 3 0...