Probes and Biopsy 13-37
3. Rotate the clamping nuts of the guided bracket on the right and left side to fix the bracket and the
transducer. Rotate the needle-type adjusting nut to the ultimate position as shown in the figure.
4. Hold the transducer. Press the biopsy needle pressure position to separate it from needle guided
V-shaped block.
5. Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.
6. Hold the transducer, and release the pressure position of the needle. Adjust the needle-type
adjusting nut manually (following the direction of the arrow). The needle moves smoothly at the
vertical direction due to its gravity.