13-56 Probes and Biopsy
Note:
Use disinfected tool to measure the distance.
Precise length data of the needle is important to the space magnetic navigation.
3.
Tap [Needle Navigation] on the touch screen to turn on the feature. Select needle type and set needle
length.
Tap [Edit Needle] and enter “Needle name” “Length” (measured in the step 1) and “needle size
(needle model)”.Tap [Add] to add vTRAX/eTRAX information in the system.You can update
vTRAX/eTRAX needle data.
You can also adjust by rotating the knob under [Needle Length] on the touch screen.
4.
Put the needle tip against the middle part of probe surface and press <Freeze>. Press <Caliper> to
enter general measurement and measure the distance between the tip and the probe surface.
5.
Put the needle body on the probe surface horizontally and press <Freeze>. Press <Caliper> to enter
general measurement and measure the distance between the needle and the probe surface.
13.3.4 Mark
Marking procedure under non-Fusion mode:
You can mark the tumor/lesion on the ultrasound image as indication for real-time biopsy.
Measure
the distance