n
v
0.014
cm
3
-------------------------
=
Appendix E
Gemini VII
E-2
239-42828-01 - Dec 2012
b.) Determine the number (
n
) of glass beads needed to equal the sample volume:
where
0.014 cm
3
= approximate volume of one bead
free
space
cm
3
2.515
g
/
cm
3
3.53
-----------------------------------------------------------------------------
mass
of
glass
beads g
=
3.
Outgas the sample in the sample tube at an appropriate temperature for an appropriate amount of
time.
4.
Install the sample tube (containing the outgassed sample) onto the analysis port and the sample
tube (containing the glass beads) onto the balance port.
5.
Specify a one-point measurement (P/Po = 0.05 to 0.1) so that the initial free-space measurement
can quickly be determined; then perform the measurement.
6.
Using the measured free space absolute value and the following relationship, determine the mass
of glass beads to remove from (or add to) the balance tube to reduce the free-space imbalance:
where
2.515 g/cm
3
= density of glass beads
3.53 = thermal correction (no units)
7.
Use a beaker of warm water to bring the balance tube to room temperature before removing it
from the balance port of the Gemini to remove (or add) glass beads. This prevents condensation
of moisture from the laboratory atmosphere onto the cold glass beads.
8.
Remove the balance tube:
• If the measured free space is negative (–), add the calculated mass of glass beads into the
balance tube.
• If the measured free space is positive (+), remove the calculated mass of glass beads from the
balance tube.
Note that the volume of one glass bead is approximately 0.014 cm
3
. Therefore, if the
measured free space is less than 0.02 cm
3
, it is unnecessary to correct the free space.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...