Schematic Shortcut Icon Available Options:
o
Cool the sample to less than 50 °C
o
Set heater power percent
Smart VacPrep Schematic Shortcuts Table (continued)
S
HOW
S
MART
V
AC
P
REP
S
TATUS
Smart VacPrep > Unit [n] > Show Status
The
Smart VacPrep Status
window allows the monitoring of degas operations for a manual degas
operation.
Field or Button
Description
Cancel
Cancels the current degas operation.
Check
Click to check the outgassing rate of the samples on the selected port.
The following process occurs:
l
The current degassing step is suspended.
l
The vacuum valves are closed and the vacuum level on the selected
ports is monitored.
l
The
Status
window indicates that the degassing operation is being
checked and displays the outgassing rate as it becomes available.
During the degas check, the
Check
button changes to
Continue
. When
Continue
is clicked, the valves open and the degassing operation
resumes.
If the outgassing rate indicated that the sample has been freed of
contaminants (minimal pressure increases), click
Skip
to advance to the
next state of the degassing operation. For example, if degassing is
checked after the setpoint is attained,
Skip
advances you to the ramping
stage.
Smart VacPrep Status Fields and Buttons Table
Show Smart VacPrep Status
Smart VacPrep Operator Manual
3 - 11
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...