Gemini VII
Appendix B
239-42828-01 - Nov 2013
B-17
4053- At least two data points must be selected for t-Plot calculations.
4054-
Fewer than two data points are inside the fitted thickness range.
4055-
A positive BET surface area was not calculated.
4056- A positive Langmuir surface area was not calculated. Please check your Lang-
muir report.
4057-
At least two data points are needed for Freundlich calculations.
Cause:
At least two points are required for the t-Plot calculations.
Action:
Edit the calculation assignments for the t-Plot report.
Cause:
At least two points must be within the fitted thickness range for the t-
Plot calculations.
Action A:
Edit the calculation assignments for the t-Plot report.
Action B:
Edit the fitted thickness range in the t-Plot report editor.
Cause:
Fewer than two points were assigned to the requested surface area
calculation in the collected data table.
Action A:
Assign more points to the surface area calculation.
Action B:
Select a different surface area in the t-Plot report editor.
Cause:
Fewer than two points were assigned to the requested surface area
calculation in the collected data table.
Action A:
Assign more points to the surface area calculation.
Action B:
Select a different surface area in the t-Plot report editor.
Cause:
Less than two data points have been selected for the Freundlich
report; at least two are required.
Action:
Select Freundlich points on the Collected Data window. If calculation
assignments are not being used, edit the Freundlich Report options,
Absolute pressure range
in the sample file.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...