Gemini VII
Appendix B
239-42828-01 - Nov 2013
B-25
6039-
Po pressure underrange.
6040-
Po pressure overrange.
6041-
Quantity adsorbed underrange.
6042-
Quantity adsorbed overrange.
Cause B:
An error occurred in the pressure measurement electronics.
Action B:
Contact a Micromeritics service representative if this error message
continues.
Cause:
A test of the Po pressure transducer showed a low reading. The
current operation will continue.
Action:
Contact a Micromeritics service representative if this error message
continues.
Cause A:
The gas pressure in the Po tube is too high. The current operation will
continue.
Action A:
Verify that the gas inlet pressure regulator setting does not exceed 15
to 18 psi (103.4 to 124.1 kPa). Contact a Micromeritics service
representative if this error message continues.
Cause B:
An error occurred in the pressure measurement electronics.
Action B:
Contact a Micromeritics service representative if this error message
continues.
Cause:
A test of the quantity adsorbed pressure transducer showed a low
reading. The current operation will continue.
Action:
Contact a Micromeritics service representative if this error message
continues.
Cause A:
The rate of adsorption by the sample exceeded the response capacity
of the analyzer, preventing normal refilling of the gas reservoirs.
Action A:
Repeat the analysis using a smaller amount of sample.
Cause B:
An error occurred in the pressure measurement electronics.
Action B:
Contact a Micromeritics service representative if this error message
continues.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...