Gemini VII
Appendix B
239-42828-01 - Nov 2013
B-21
4135-
HOA file [filename] does not exist.
4136-
HOA [filename] is corrupt.
4400-
The computer does not have the communications port specified for the Smart-
Prep. Cannot initialize.
4401-
The communications port specified for the SmartPrep is already in use. Cannot
initialize.
4402-
The communications port specified for the SmartPrep cannot be accessed.
Cannot initialize.
4403-
Cannot communicate with SmartPrep Unit [n] - S/N: [nn].
4404-
The application version of the SmartPrep Unit [n] - S/N: [nn] is invalid.
Cause:
The sample file in the
Heat of Adsorption
report list does not exist.
Action:
Go to
Report > Heat of Adsorption
. Click
Add Samples,
then select
the sample file.
Cause:
The sample file in the
Heat of Adsorption
report is corrupt.
Action:
Go to
Report > Heat of Adsorption
. Select the correct sample file,
then click
Remove Sample
. Rerun the
Heat of Adsorption
report.
Cause:
The communications port associated with this unit was not valid.
Action:
Run the setup program and set up the unit on a valid port.
Cause:
The communications port associated with this instrument is in use by
some other program in the system.
Action:
Close the other program to release the port. Restart the analysis
application.
Cause:
The communications port associated with this unit was not valid.
Action:
Run the setup program and set up the unit on a valid port.
Cause:
The controller software running on the designated instrument is
invalid.
Action:
Use the SmartPrep setup program to download the proper controller
software to the instrument, or if unavailable, contact a Micromeritics
service representative.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...