3 Operate the Smart VacPrep
10. Install the safety shield.
11. Press the
Select
button for the port to be degassed. A blue LED is lit.
12. Press the
Start
button on the Smart VacPrep panel to start the degas process for all ports that
have a blue LED lit.
13. When degassing is complete and the sample has cooled, the sample can be transferred to an ana-
lysis port.
L
OAD
Q
UICK
S
TART
D
EGAS
C
ONDITIONS INTO A
S
AMPLE
F
ILE
QuickStart Degas Conditions can be copied into a sample file. Reference the Degas Conditions section
of the analyzer Operator Manual.
Field or Button
Description
Close
Closes the active window.
Copy QuickStart Degas
Conditions from the
Smart VacPrep Unit
Use to copy the degas conditions settings from the selected Smart
VacPrep unit and port.
Degas Conditions
Use to browse for a .DEG file that contains degas condition parameters
to be used in the analysis.
Heating Phase
This option is applicable when degassing with either a Smart VacPrep or
a SmartPrep.
Enter up to five stages of degas conditions.
l
Temperature.
Degas temperature.
l
Temperature Ramp Rate.
Rate at which the temperature is to
change.
l
Time.
Amount of time to heat the sample.
Use to modify the table contents.
l
Insert.
Inserts one row above the selected row.
l
Delete.
Deletes the selected row.
l
Clear.
Clears all table entries and displays only one default value.
l
Append.
Inserts one row at the end of the table.
Degas Conditions Fields and Buttons Table
3 - 4
Smart VacPrep Operator Manual
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...