8 Error Messages
4412
Error calibrating the servo.
Cause:
Calibration results are out of range.
Action:
Follow standard calibration procedures and try again. If the problem persists, contact
your Micromeritics service representative.
4413
Error overheating on port [0]. Current = [TEMPC0] [TEMPC-U], Target
= [TEMPC0] [TEMPC-U], Limit = [TEMPC0] [TEMPC-U].
Cause:
The temperature of the indicated mantle exceeded the maximum allowed value.
Action:
Ensure the power and thermocouple connectors for the mantle are properly installed. If
they are, the mantle may be defective; contact your Micromeritics service rep-
resentative.
4414
Error thermocouple unplugged on port [0]. Target = [TEMPC0]
[TEMPC-U]."
Cause:
The thermocouple is unplugged or has malfunctioned.
Action:
Ensure the thermocouple is plugged in. If the problem persists, contact your Micro-
meritics service representative.
4420
Communications error.
Cause:
The application failed to connect to the Smart VacPrep.
Action:
Ensure the unit is powered and properly connected to the network specified during install-
ation. Check the power cable, power switch, and Ethernet cable, then reconnect to the
Smart VacPrep either through the Smart VacPrep menu for this instrument or by restart-
ing the application. If the problem persists, contact your Micromeritics service rep-
resentative.
4421
Smart VacPrep S/N %s is busy and could not be removed.
Cause:
The Smart VacPrep cannot be removed because it is currently performing an operation.
Action:
Wait until the Smart VacPrep completes the current operation and try again.
8 - 2
Smart VacPrep Operator Manual
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...