Appendix B
Gemini VII
B-20
239-42828-01 - Nov 2013
4073-
The analysis gas [name] does not match the model gas [name].
4074-
The analysis temperature [nn] does not match the model temperature [nn].
4075-
The models cannot be located in the models folder. Reinstall the software.
4077-
Cannot get surface area for: [file name]
.
4078-
Slope and Y-Intercept cannot be determined from the selected points.
4079-
Points found with negative pressure values will not appear on the graph.
Cause:
The model assumes a specific gas and the sample file uses a different
one.
Action:
Select a model that assumes the same gas.
Cause:
The temperature for the selected model did not match the analysis
temperature.
Action:
Select a different model.
Cause:
The models could not be located. They may have been inadvertently
deleted or moved.
Action:
Reinstall the software.
Cause:
The Isotherm report for the named overlay file has
Per gram
selected
for the
Volume Adsorbed
and the Isotherm report for the primary file
has a surface area option selected for the
Volume Adsorbed
.
Action A:
Edit the Isotherm report for the named overlay file and select a
surface area option for
Volume Adsorbed
.
Action B:
Click
Overlays
on the Report options window of the primary file and
remove the named overlay file from the list.
Cause:
The Langmuir report cannot be generated from the selected points.
Action:
Edit the calculation assignments in the Langmuir column on the
Collected Data window.
Cause:
Collected data contains negative pressure values.
Action:
Negative pressure points are automatically removed from isotherm
graphs. To display negative pressure values appear on the graph:
Open the sample file and click the
Collected Data
tab. Deselect
Use
calculation assignments
and then select the points with negative
pressures as Outliers.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...