Appendix B
Gemini VII
B-26
239-42828-01 - Nov 2013
6043-
Evacuation timeout.
6044-
Dosing timeout.
6045-
Elevator timeout.
6048-
Po fill timeout.
Cause:
The analyzer failed to reach a suitable vacuum level within the
specified evacuation time.
Action A:
Remove and reattach the sample and balance tubes, ensuring that both
are securely in the port.
Action B:
Verify that the vacuum pump is operating properly.
Action C:
If actions A and B don’t correct the problem, degas the sample again
and repeat the analysis.
Action D:
Contact a Micromeritics service representative if this error message
continues.
Cause A:
The adsorbate gas supply was depleted by the analyzer during
analysis or termination.
Action A:
Verify that the inlet valve on the gas bottle is open and that the gas
pressure regulator setting is between 15 and 18 psig (103.4 and 124.1
kPa). Check the adsorbate gas bottle to see if it is empty; replace if
necessary.
Cause B:
Sample has adsorbed the maximum amount of gas available.
Action B:
If additional data points are required, repeat the analysis with a
smaller amount of sample (to avoid exceeding 100 cm
3
STP total
volume adsorbed).
Cause:
The elevator was unable to reach the raised or lowered position.
Action:
Check for obstructions. Contact a Micromeritics service
representative if this error message continues.
Cause:
The adsorbate gas supply was depleted during and analysis or a Po
measurement.
Action:
Verify that the inlet valve on the gas bottle is open and that the gas
pressure regulator setting is between 15 and 18 psig (103.4 and 124.1
kPa). Check the adsorbate gas bottle to see if it is empty; replace if
necessary.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...