3 Operate the Smart VacPrep
Degas conditions remain the same from degas to degas unless the
Degas Conditions
file is changed or
replaced by the operator. A manual evacuation can be performed on any idle degas port.
The six ports are represented by row numbers. Ports that are busy are grayed out and disabled.
1. Click
Browse
to search for a sample file or a degas conditions file. Select the file, then click
Open
. Do this for each Smart VacPrep port to be used.
2. If using a Check Seal, verify that the Check Seal opener is installed in the Smart VacPrep port.
Follow the instructions included with the Check Seal for instructions on inserting the opener.
3. Load the sample into the sample tube. If using a Check Seal or TranSeal, insert it into the sample
tube.
4. Load the sample on the Smart VacPrep sample port.
5. Attach the heating mantle using the metal clip. If a straight wall tube is used, heating mantles
with elastic cords are available. Refer to the Parts and Accessories section of the analyzer oper-
ator manual.
6. Install the safety shield.
7. Click
Start
on the
Start Degas
window to begin the degas process for the selected Smart
VacPrep ports.
8. Click
Start
to begin the degas process for the selected Smart VacPrep ports.
9. Allow the samples to cool before transferring them to the analysis ports to start the analysis.
Field or Button
Description
Browse
Searches for a file. Select a file from the
Name
column or from the
library, then click
Open
. Alternatively, double click the file name to
open (or import) the file.
Cancel
Discards any changes or cancels the current process.
Clear
Clears the field.
Sample
Displays the path and file name of the selected .SMP file.
Start
Starts the degas operation for the selected ports.
Smart VacPrep Start Degas Fields and Buttons Table
3 - 6
Smart VacPrep Operator Manual
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...