Analysis Conditions Files
Gemini VII
5-16
239-42828-01 - Dec 2012
p
o
and T button
(continued)
•
Entered
- select to manually enter the Po in the text box.
•
Most recent measured
- select to use the most recently measured
saturation pressure.
•
Measure Po for each point (models 2390p and 2390t only)
- select if
Po is to be measured for each point.
•
Bath temperature
- calculates Po using the entered bath temperature.
Analysis Method
button
Select the method of analysis.
•
Equilibrate
- select if the analysis method is to be equilibrated and
enter the equilibration time in seconds between successive pressure
readings during equilibration. With this method, the analyzer increases
pressure in steps to the next specified level for data collection. It
maintains this pressure and monitors volumes (or quantities) adsorbed.
When the rate of adsorption (volume adsorbed) falls below 0.1%,
equilibration is assumed to have occurred. The instrument then
introduces the next adsorbate gas dose.
•
Scan
- applies to adsorption only. Select if the analysis method is to be
scanned and enter the minimum equilibration in minutes. This method
allows the adsorbate gas pressure to be raised continuously at a
constant rate based on the pressures to be achieved and the length of the
analysis.
Cancel button
Close button
OK button
Save button
Table buttons
Common Fields and Buttons - File Menu Options
.
Field or Button
Description (continued)
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...