Appendix B
Gemini VII
B-8
239-42828-01 - Nov 2013
2510-
[sample file] Error parsing reports from file [name]. Reports cannot be
produced.
2511-
Print job [name] has been cancelled due to insufficient disk space. Delete
unnecessary files and restart the report.
2512-
Print job [name] cancelled.
2513-
Unable to read the calibration file [file name].
2514-
Unable to write the calibration file [file name].
Cause A:
One or more data entry fields in the sample file may contain an
invalid character (such as a single quote or double quotes).
Action A:
Review the data entry fields (for example, the
Sample
field) and
remove the invalid character.
Cause B:
The system was unable to create the usual temporary files during the
report, possibly due to insufficient disk space.
Action B:
Check the space available on the hard disk.
Cause C:
An internal processing error occurred.
Action C:
Contact a Micromeritics service representative if this error message
continues.
Cause:
The disk drive does not have required space for the temporary file.
Action:
Delete unnecessary files from the disk. At least five megabytes of free
space is required for normal operation.
Cause:
The print job was canceled by the operator.
Action:
None required.
Cause:
An invalid calibration file was selected or one that cannot be read.
Action:
Ensure the media containing the calibration file has no problems.
Cause:
An attempt to save calibration data has failed due to possible media
problems.
Action A:
Ensure the destination location has no problems.
Action B:
Choose an alternate media to save the calibration data.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...