Gemini VII
Appendix B
239-42828-01 - Nov 2013
B-13
4036-
Fewer than 2 points available for Horvath-Kawazoe calculations.
4037-
Computations failed while processing the primary data set. No reports will be
produced.
4038-
Fewer than 2 points available for the Langmuir Qm computation. Cheng/Yang
correction will not be applied.
Cause:
At least two data points must be selected for inclusion in the Horvath-
Kawazoe analysis on the Collected Data window. No report will be
produced.
Action:
From the Collected Data window, select the data points to be
analyzed.
Cause:
The preparation of data for reporting could not be successfully
completed. No Horvath-Kawazoe reports will be produced. This
message will always be preceded with another one containing
additional information.
Action:
Refer to the number of the error message which preceded this one for
an explanation.
Cause:
The Cheng/Yang correction to the Horvath-Kawazoe equation
requires the value of the monolayer volume as calculated from the
Langmuir equation. The isotherm must include at least two points
above 0.02 relative pressure for the Langmuir equation to be applied.
Action:
The analysis will be performed without the Cheng/Yang correction.
Deselect Apply Cheng/Yang correction on the Horvath-Kawazoe
Report Options window to prevent this message from appearing on
future reports.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...