82
Perform the visual inspection of the appliance / equipment.
Apply appropriate ticker(s) to items of inspection.
End inspection.
Save results (optional).
Figure 6.15: Examples of Visual inspection results
6.2.2 Continuity
Figure 6.16: Continuity test menu
Test results / sub-results
R
............... Resistance
ΔU
………..Voltage drop scaled to 10 A
Test parameters
Output
Output connections [4-wire, P/S - PE, MS_PE - IEC_PE]
I out
Test current [0.2 A, 4 A, 10 A, 25 A]
Duration
Duration
[Off, 2 s ... 180 s]
ΔU test
Enables
ΔU test* [On, Off]
Wire cross-section
Wire cross-
section for ΔU test ** [0.5 mm
2
…
6mm
2
]
Test limits
H limit(R)
High limit (R)
[Off, 0.01 Ω ... 9 Ω, Custom]
Limit (ΔU)
Limit (ΔU)** [calculated value]
* Applicable only at test current 10 A.
** Applicable only at
ΔU test set to On