MI 3325 MultiServicerXD
22
The measured leakage (Idiff, Ipe, Itouch) current was
higher than 20 mA. Measurement was aborted. Press
OK to continue.
The load current higher than 16 A is detected.
Measurement is aborted. Press OK to continue.
The average load current higher than 10 A over the last
5 min test interval is detected. Measurement is
stopped. Recovery period for cooling is required before
proceeding with tests!
Press OK to continue.
The polarity pre-test of the cable / PRCD has failed.
Press OK to continue.
Yellow screen and activated buzzer sound!
Hazardous live voltage on TP1-PE input is present
before running RCD tests, Rpe, Z loop, Zs rcd, Z auto,
Voltage drop ΔU and Auto Sequences
®
.
Warning!
Immediately disconnect supply from tested
installation / equipment and check and
correct PE wiring!