MI 3325 MultiServicerXD
204
No notifications mode
Instrument skips pre-test warnings (see chapter
information).
Parameters
State
On
– enables No notifications mode
Off
– disables No notifications mode
Appliance info
Instrument enables to automatically add the appliance name to the Auto Sequence
®
.
Parameters
Repeat Setting
Repeat:
The same Appliance ID will be offered each time if the
same Auto test is carried out successively in a loop.
Increment:
A four digit number will be added to the Appliance ID
and incremented each time if the same Auto
Sequence® is carried out successively in a loop.
Appliance type
Selects the type of the appliance (Appliance, Appliance_FD, Welding
equip., Welding equip. FD)
Default Appliance ID Enter default Appliance ID
Appliance name
Enter Appliance name.
Options:
Editable
– allows Appliance name to be modified while running
Auto Sequence®. Menu with a list of Appliance names and possibility
to enter custom Appliance name is offered within the test.
Not editable
– Default Appliance name is used. Appliance name
cannot be modified while running Auto Sequence®.
Retest period
Retest period in months.
Options:
Editable
– allows Retest period to be modified while running Auto
Sequence®. Numeric keypad for entering custom Retest period is
offered within the test.
Not editable
– Default Retest period is used. Retest period cannot
be modified while running Auto Sequence®.
Note
This flow command is active only if Auto Sequence
®
is started from the Auto Sequences
®
Main menu.
Inspection Expert mode
If Inspection Expert mode flow command is set, the Visual inspection screen and Functional
inspection screen within Auto Sequence
®
are displayed for 1 second and an overall PASS is
automatically applied at the end of test. In between, the automatic procedure can be stopped
and statuses can be applied manually.
Inspection Expert mode is disabled by default.