MI 3325 MultiServicerXD
21
Select
YES
to proceed with or
NO
to cancel
measurement.
Warning for proper connection in PE_conductor
measurements. Connection of PRCD’s plug must be
changed in order to proceed.
Select
YES
to retry after reconnection with or
NO
to
cancel measurement.
Warning for improper supply voltage condition. If
pressing OK instrument will continue to work in a
limited mode (measurements are disabled).
In pre-test an external voltage between C1/P1 and
C2/P2 terminals was detected. The measurement was
cancelled. Press
OK
to continue.
In pre-test a too high external voltage was detected
between P and PE terminals. The measurement was
cancelled. Press OK to continue.
In pre-test a too high external voltage was detected
between IEC test socket and PE terminal. The
measurement was cancelled. Press OK to continue.
In pre-test a too high external voltage was detected
between Current Clamp and PE terminals. The
measurement was cancelled. Press OK to continue.
In pre-test a possible high leakage current was
detected. It is likely that a dangerous leakage current
(higher than 3.5 mA) will flow after applying power to
the device under test.
Select
YES
to proceed with or
NO
to cancel
measurement.