MI 3325 MultiServicerXD
99
Figure 6.49: Examples of Touch Leakage measurement results
6.2.11 Welding Circuit Leakage I leak (W-PE)
Figure 6.50: Welding Circuit Leakage test menu
Test results / sub-results
Iw
............. Welding circuit Leakage current
Test parameters
Duration
Duration
[Off, 2 s
… 180 s]
Change
Change status
[YES, NO]
YES: The instrument measures leakage current in two sequential steps
with delay* in between. The phase voltage is firstly applied to the
right live output of the mains test socket and secondly to the left
live output of the mains test socket.
NO: The phase voltage is applied only to the right live output of the
mains test socket.
Delay
*Delay time
[0.2 s … 5 s]
Test limits
Limit (Iw)
Limit(Iw) [Off, 3.5 mA
… 10 mA ]
Test circuit, I leak (W-PE) measurement procedure
Select the
I leak (W-PE)
function.
Set test parameters / limits.
Connect METREL 3-phase adapter (A 1422) to the instrument*.
Connect device under test to the 3-phase adapter*.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
*For more information refer to chapter
Measurements according to IEC/ EN 60974-4
– Welding
circuit leakage current in 3-phase adapter user manual.