MI 3325 MultiServicerXD
138
Test circuit
Figure 6.113: HV AC programmable test
HV AC programmable test procedure
Prepare test setup as mentioned in IMPORTANT SAFETY NOTES above.
Select the
HV AC
programmable
function.
Set test parameters / limits.
Connect HV test leads to HV terminals on the instrument.
Connect HV test leads to device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 6.114: Examples of HV AC programmable test results
Note:
First HV measurement after power on the instrument (if password protection is enabled)
or first HV measurement after enabling or changing password require entering password
for enabling HV test. Refer to chapter
4.7.4 Change password for HV functions
more information.