MI 3325 MultiServicerXD
33
Figure 4.7: Change password menu
Notes:
Default password is 0000.
An empty entry disables the password.
If password is lost, entry 4648 resets password to default.
4.7.5 Initial Settings
In this menu internal Bluetooth module can be initialized and the instrument settings,
measurement parameters and limits can be set to initial (factory) values.
Figure 4.8: Initial settings menu
Warning!
Following customized settings will be lost when setting the instruments to initial settings:
Measurement limits and parameters
Global parameters, System settings and Devices in General settings menu
Opened Workspace and Auto Sequence® group will be deselected.
User will be signed out.
Note:
Following customized settings will stay:
Profile settings
Data in memory (Data in Memory organizer, Workspaces, Auto Sequence
®
groups and
Auto Sequences
®
)
User accounts
Password for HV functions