MI 3325 MultiServicerXD
103
6.2.14 Power
Figure 6.57: Power measurement menu
Test results / sub-results
P
............... Active power
S
............... Apparent power
Q
.............. Reactive power
PF
............. Power factor
THDu
........ Total harmonic distortion
– voltage
THD
i ......... Total harmonic distortion
– current
Cos Φ
....... cosines
Φ
I
................ Load current
U
............... Voltage
Test parameters
Duration
Duration
[Off, 2 s
… 180 s]
Test limits
H limit(P)
High Limit (P) [Off, 10 W
… 3.50 kW, Custom]
L limit(P)
Low Limit (P) [Off, 10 W
… 3.50 kW, Custom]
Test circuit
Figure 6.58: Power measurement
Power measurement procedure
Select the
Power
function.
Set test parameters / limits.
Connect device under test to the instrument (see test circuits above).
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).