PAT Appliances & Machines Safety Trainer
AD 3
– MI 3399
243
5.5 Portable appliance of class 1
– Extension drum
Simulates
Error value
Error off Error on
Simulation between contacts
Measurement taken
Error off Error on
Error 1
Loose contact of PE conductor
Error value
0,02 Ω
0,84 Ω
PE connection on output socket
(1) and PE connection on mains
socket (2)
Earth Bond
0,02 Ω
0,84 Ω
Error 2
Insulation fault
Error value
>20 MΩ 106 kΩ
L and PE pins on socket (1)
Insulation 500 V DC
>20 MΩ 0,106 MΩ
Leakage 230 V,
50/60 Hz
0,00 mA 2,17 mA
Subleakage
0,00 mA 2,17 mA
Error 3
Insulation fault
Error value
30 MΩ
238 kΩ
Isolated accessible metal part (3)
and L, pin on socket (2)
Insulation 500V DC
(probe)
13 MΩ
0,237 MΩ
Touch Leakage 230 V
AC
0,00 mA 0,97 mA
Subleakage (probe)
0,00 mA 0,97 mA
Table 5-4: Description of simulated faults / relevant tests
Содержание MI 3399
Страница 27: ...Electrical Installation Safety Trainer AD 1 MI 3399 27...
Страница 28: ...Electrical Installation Safety Trainer AD 1 MI 3399 28 AD 1 MI 3399 EIS Electrical Installation Safety Trainer...
Страница 46: ...Electrical Installation Safety Trainer AD 1 MI 3399 46 3 Exit settings...
Страница 74: ...Electrical Installation Safety Trainer AD 1 MI 3399 74...
Страница 122: ...Power and Voltage Quality Trainer AD 2 MI 3399 122 AD 2 MI 3399 PQA Power and Voltage Quality Trainer...
Страница 197: ...Power and Voltage Quality Trainer AD 2 MI 3399 197 Figure 4 89 Transient Chart view...
Страница 204: ...PAT Appliances Machines Safety Trainer AD 3 MI 3399 204...
Страница 205: ...PAT Appliances Machines Safety Trainer AD 3 MI 3399 205 AD 3 MI 3399 PAT Appliances Machines Safety Trainer...
Страница 249: ...PV Photovoltaic Systems Trainer AD 4 MI 3399 249...
Страница 250: ...PV Photovoltaic Systems Trainer AD 4 MI 3399 250 AD 4 MI 3399 PV Photovoltaic Systems Trainer...