Power and Voltage Quality Trainer
AD 2
– MI 3399
178
Figure 4-58: Expanded Power & Energy node
4.6 Exercise 6
– Harmonics measurement
In this exercise, we will show how instrument should be setup in order to measure and
analyze harmonics. Before start, please set the switch S5 in the Application Trainer to
ON as shown in the table below. The remaining switches should be set to OFF.
Switch
Description
State
S5
Harmonics
ON
The Application Trainer will simulate an electrical system with the following
parameters:
Fundamental voltage: 230 V
Fundamental current: 1000 A
Load type: Inductive
Load character: Load (Consumption of electrical power)
Frequency: 50/60 Hz
Voltage harmonics:
o
3
rd
harmonic:5%,
o
5
th
harmonic: 5%,
o
7
th
harmonic: 5%
Current harmonics:
o
3
rd
harmonic:15%,
Содержание MI 3399
Страница 27: ...Electrical Installation Safety Trainer AD 1 MI 3399 27...
Страница 28: ...Electrical Installation Safety Trainer AD 1 MI 3399 28 AD 1 MI 3399 EIS Electrical Installation Safety Trainer...
Страница 46: ...Electrical Installation Safety Trainer AD 1 MI 3399 46 3 Exit settings...
Страница 74: ...Electrical Installation Safety Trainer AD 1 MI 3399 74...
Страница 122: ...Power and Voltage Quality Trainer AD 2 MI 3399 122 AD 2 MI 3399 PQA Power and Voltage Quality Trainer...
Страница 197: ...Power and Voltage Quality Trainer AD 2 MI 3399 197 Figure 4 89 Transient Chart view...
Страница 204: ...PAT Appliances Machines Safety Trainer AD 3 MI 3399 204...
Страница 205: ...PAT Appliances Machines Safety Trainer AD 3 MI 3399 205 AD 3 MI 3399 PAT Appliances Machines Safety Trainer...
Страница 249: ...PV Photovoltaic Systems Trainer AD 4 MI 3399 249...
Страница 250: ...PV Photovoltaic Systems Trainer AD 4 MI 3399 250 AD 4 MI 3399 PV Photovoltaic Systems Trainer...