Power and Voltage Quality Trainer
AD 2
– MI 3399
137
Figure 3-4: Setting Nominal voltage on the instrument
3.
Exit the nominal voltage setup by pressing once the
key.
4.
Select Phase Current Clamps menu and press the
key
5.
Select A1033 Clamps and press the
key
Figure 3-5: Current clamp selection
6.
Select 100% measuring range and press the
key.
Figure 3-6: Setting Current clamps parameters on the instrument
7.
Select Neutral Current Clamps and repeat procedure above (steps 5 and 6).
8.
Select Connection menu and press the
key
ENTER
ENTER
ENTER