7 | Operation
92
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
exISO max
Upper alarm limit for the expiratory isoflurane concentration
exISO min
Lower alarm limit for the expiratory isoflurane concentration
inSEV max
Upper alarm limit for the inspiratory sevoflurane concentration
inSEV min
Lower alarm limit for the inspiratory sevoflurane concentration
exSEV max
Upper alarm limit for the expiratory sevoflurane concentration
exSEV min
Lower alarm limit for the expiratory sevoflurane concentration
7.3.7.3
Other alarms
RR max_spontan
Upper alarm limit for the measured spontaneous respiratory rate
RR min_spontan
Lower alarm limit for the measured spontaneous respiratory rate
Pmean max
Upper alarm limit for the measured mean airway pressure
Pmean min
Lower alarm limit for the measured mean airway pressure
Pplateau max
Upper alarm limit for the measured plateau pressure
HR (pulse) max
Upper alarm limit for the measured heart rate
HR (pulse) min
Lower alarm limit for the measured heart rate
SpO
2
max
Upper alarm limit for the measured SpO
2
concentration
SpO
2
min
Lower alarm limit for the measured SpO
2
concentration
7.3.7.4
Alarm limits for non-invasive ventilation
In an effort to prevent false alarms, to reduce the stress level for the ventilated
patient due to alarm sounds and to maintain reliable monitoring, the alarm limits
need to be adapted to the patient’s individual situation.
For the alarms listed below, a delay can be defined, which should take the clinical
situation into account. This will cause a delay of the audible and visual alarms:
Alarm delay
Default value
Adjustment range
T VT max
5 s
0–15 s
T VT min
5 s
0–15 s
T P min
10 s
0–30 s
T Backup ventilation
30 s
5–120 s
T Apnoea alarm
20 s
5–60 s
T Leakage high
5 s
0–15 s
T Disconnection
10 s
0–30 s
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...