A 5 | Ventilation modes and parameters
110
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
The Pmax and Pmin parameters determine a range in which the ventilator is
allowed to automatically readjust the necessary pressure to achieve the desired
tidal volume.
In contrast to conventional SIMV ventilation, the user defines a minimum minute
volume by setting the respiratory rate and tidal volume. No mandatory breaths
are initiated if the minute volume can be achieved by the patient's spontaneous
breathing.
The ventilator continuously monitors the patient's respiratory activity. If the min-
ute volume drops below the preselected value, a mandatory breath is initiated. If
spontaneous breathing fails completely, the patient is ventilated with the selected
mandatory respiratory rate.
A 5.5.10
PC-SIMV
Pressure Controlled - Synchronised Intermittent Mandatory Ventilation. Synchro
-
nised, pressure-controlled ventilation with the possibility of spontaneous breath-
ing between the controlled breaths.
This mixed form of ventilation combines a synchronised, pressure-controlled ven
-
tilation mode to ensure a specified minute volume with the possibility of achieving
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...