9 | Alarms and troubleshooting
134
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#238
3
Paux1 pres
-
sure error
Paux1 pres
-
sure sensor
error
Information:
- The Paux1 function is disabled
- Other device functions are not affected
Ventilator is operating safely.
- Contact an authorised service agent.
#239
3
Insp. pres-
sure error
Inspiration
pressure
sensor error
Information:
- The inspiratory pressure sensor is defective
- Patient pressure is measured and controlled only by
the expiratory pressure sensor
Ventilator is operating safely.
Contact an authorised service agent.
7
Information:
- Safe operation cannot be guaranteed
Do immediately:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
Contact an authorised service agent.
#240
3
Exp. flow
error
Expiratory
flow sensor
error
Information:
- A redundant sensor is defective
- The accuracy of low readings can be affected
Ventilator is operating safely.
7
Information:
- The expiratory flow sensor system is defective
- Tidal and minute volume cannot be measured
- Ventilator now operates in emergency ventilation
mode
Decide: If emergency ventilation mode is a hazard:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
Contact an authorised service agent.
#241
3
Paux2 pres
-
sure error
Paux2 pres
-
sure sensor
error
Information:
- The Paux2 function is disabled
- Other device functions are not affected
Ventilator is operating safely.
Contact an authorised service agent.
#243
3
O2 flow
error
O2 flow
sensor error
Information:
- A redundant sensor is defective
- The accuracy of low readings can be affected
Ventilator is operating safely.
#243
7
O2 flow
error
O2 flow
sensor error
Information:
- The O2 flow sensor system is defective.
- Ventilation is now only possible with air
Monitor:
- O2 concentration
Decide: If ventilation with low oxygen is a hazard:
- Disconnect the patient from this ventilator
- Put this ventilator out of service
- Continue ventilation with another ventilator
- Contact an authorised service agent.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...