Operation | 7
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
97
Remark
The remaining time can vary depending on the load. The informa-
tion is updated in intervals of one minute.
7.4.4
System test
Information about the most recent system test is displayed. Furthermore, you can
perform a system test by pressing the “Start system test” button.
7.4.5
Settings
The “Settings” menu is available in the standby mode.
To change a setting, touch the appropriate button, set the new value with the
selector, and confirm the new value with the Enter button.
7.4.5.1
Date / Time
Manual setting of date and time. This function is not available during ventilation
when the field is dimmed.
7.4.5.2
Display brightness
The device offers three options for adjusting the screen brightness:
•
Day mode
•
Night mode
•
Night screen
Using the selector, you can preset a brightness level for the day mode and for the
night mode: the day brightness and the night brightness. By selecting the day or
night mode you activate the corresponding brightness level.
If an alarm of medium priority or above occurs, the display brightness is automat-
ically set to the level of the day brightness.
In addition, the “Night screen on” option activates a dimmed screen display with
a reduced number of data. This option can be selected only when the device is in
the night mode.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...