Alarms and troubleshooting | 9
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
161
Paux Alarms
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#751
1
Paux Error
The power
supply of the
Paux pump is
defective
Information:
- Any Paux function is only available when oxygen
is connected.
If the alarm recurs or persists:
Contact an authorised service agent.
5
Information:
- Any Paux function is only available when oxygen
is connected.
Do immediately:
- Connect ventilator to oxygen gas supply
If the alarm recurs or persists:
Contact an authorised service agent.
#752
1
Paux Error
The current
consumption
of the Paux
pump is
outside the
permitted
range
Information:
- Any Paux function is only available when oxygen
is connected.
If the alarm recurs or persists:
- Contact an authorised service agent.
#752
5
Paux Error
The current
consumption
of the Paux
pump is
outside the
permitted
range
Information:
- Any Paux function is only available when oxygen
is connected.
Do immediately:
- Connect ventilator to oxygen gas supply
If the alarm recurs or persists:
- Contact an authorised service agent.
#753
5
Paux error
Paux supply
pressure is
high.
Information:
- All Paux functions are disabled
If the alarm recurs or persists:
- Contact an authorised service agent.
#754
5
Paux error
Paux supply
pressure is
low.
Information:
- All Paux functions are disabled
If the alarm recurs or persists:
- Contact an authorised service agent.
9.5
Error messages during power up test
Remark
Please note down the displayed error codes for service purposes.
Code
Error message
Corrective measure
#201
to
#269
Pressure, flow or voltage meas
-
urement error
(sensor problem)
Do not use ventilator, switch OFF,
contact authorised service agent
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...