Operation | 7
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
87
7.3.1
Set ventilation mode
A narrow coloured bar at the top of the screen identifies the active ventilation
mode.
NIV (non-invasive)
Depending on selected screen colour:
green
or
blue
IV (invasive)
grey
Backup ventilation
yellow
Resuscitation (or CPR) mode
red
The backup mode is indicated in the backup mode window.
Open the menu for selection of the ventilation mode with the “Ventilation” but-
ton in the button bar. Select one of the preconfigured ventilation modes and adapt
the ventilation parameters to the patient's needs (also refer to chapter 7.3.3“Set
ventilation parameters”).
7.3.2
Change ventilation mode
To switch between ventilation modes during ventilation, select the “Ventilation”
button in the button bar. Activate the new ventilation mode in the ventilation
mode bar, then press the “Change Mode: Manual” button or the “Change Mode:
Wizard” button, and confirm the change with the Enter button.
A wizard is available for switching between volume-controlled and pressure-con-
trolled ventilation modes. On the basis of the measured values, the assistant rec-
ommends a pressure or a volume setting.
The plateau pressure values measured last, for example, are used to set Pinsp.
With the arrow button
you display the advanced parameters. The number of
advanced parameters available depends on the configuration settings.
Remark
The “Change Mode: Wizard” function is not available when the
current measured values cannot be used as the basis for an auto-
matic switchover. The button is inactive.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...