Alarms and troubleshooting | 9
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
111
9.2
Alarm system
The intensive care ventilator is equipped with alarms to support patient safety.
The functions of the alarm system are automatically verified as part of the power
up test and the system test.
When an alarm condition occurs, it is indicated in the alarm window and it also
triggers the audible and visual signalling devices.
The system classifies the alarms in 7 different priority levels which can also esca
-
late and de-escalate.
The alarm messages in the alarm window are sorted according to activity, priority
and time of occurrence. If more than 4 alarms occur at the same time, the 4 alarm
messages with the greatest importance will be shown.
The audible and visual alarms from the signalling devices are always those of the
active alarm with the highest priority.
After switching between invasive and non-invasive ventilation or between ventila-
tion modes, check the alarm limits and ventilation settings and adjust them so that
the ventilation can be fully monitored. Extreme alarm limit settings may render
the alarm system useless.
The user is required to review the alarm limits to determine their appropriate-
ness for the individual ventilation situations and to adjust them if necessary.
Warning
Switch off alarm limits only if the missing alarm does not consti-
tute a safety hazard for the patient!
9.2.1
Alarm messages
When you touch an alarm message in the alarm window, the window expands
and shows information on the cause and possible corrective measures. If an alarm
becomes inactive, the audible and visual signalling devices are also deactivated. The
alarm message text, however, remains visible in the alarm window for informa
-
tional purposes until it is acknowledged with the alarm pause button (
A
).
By touching the respective arrows, the alarm window can be closed horizontally
or vertically. In this case, the alarm with the highest priority and the number of
additional alarms will be displayed. After 20 seconds all alarms are displayed again
in full size.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...