Operation | 7
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
89
Remark
The adjustment ranges of the ventilation parameters can limit
each other and may vary, depending on the situation.
The selector indicates the parameter which is limiting an adjust-
ment range.
To cancel the entry, press the button of the selected parameter once again.
7.3.4
Trigger
The trigger function synchronises the ventilator and the patient. Depending on
the selected ventilation mode and patient activities, mandatory ventilation strokes
can be initiated via a flow or a pressure trigger, or the pressure support PS can
be activated to enable deeper spontaneous breaths. The ventilator detects the
patient's attempts to breathe in via the trigger, and the trigger sensitivity can be
individually adapted to the patient. The lower the set value the more sensitive the
ventilator is to patient's attempts at inhaling. At very low sensitivities, the elasticity
of the ventilation tubes could cause a trigger signal (auto-triggering).
Inspirations initiated by the trigger variable are coloured on the pressure curve.
Different colours are assigned to mandatory and spontaneous triggers for differ-
entiation.
If the triggered breaths are not coloured, the function may not have been activat-
ed. Change the configuration accordingly (to open the menu, tap and keep your
finger on the curve) or contact your service technician.
Switching between flow and pressure trigger is done with the selector at the
minimum trigger value or at OFF. The indication “Flow” and “Press.” is displayed
as appropriate. The desired trigger function is available after selection and con-
firmation.
Remark
When the trigger function is switched off, no pressure support is
available in any of the modes to assist spontaneous breathing.
7.3.5
Backup ventilation
A backup ventilation mode, which completely takes over the patient's ventilation
in case of apnoea, is associated with each ventilation mode.
Tap the “Backup” button to configure the backup ventilation parameters.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...