13 | Default parameters
208
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
13.1
Default parameters (factory settings)
13.1.1
Default parameters
Patient category
Adults
Children
Ventilation mode at
start-up
BiLevel
BiLevel
Tube compensation
Comp.: 100%
Diam.: 7.5 mm
Insp.: On
Exp.:
Off
Comp.: 50%
Diam.: 6.0 mm
Insp.: Off
Exp.:
Off
O
2
50%
50%
P insp
12 mbar + PEEP
9 mbar + PEEP
PEEP
5 mbar
5 mbar
Insp. Ramp
0.10 s
0.10 s
Exp Ramp mand
OFF
OFF
Exp Ramp spont
OFF
OFF
Rate
12 /min
20 /min
T insp
1.5 s
1.0 s
Vt
400 ml
150 ml
I-Flow
40 l/min
25 l/min
T Plateau
0.5 s
0.5 s
Trigger (Flow)
5.0 l/min
3.0 l/min
Exp. trigger
0%
0%
PS
10 mbar + PEEP
9 mbar + PEEP
PS Endflow
25%
25%
PS Ramp
0.10 s
0.10 s
PS TImax
4.0 s
3.0 s
T_exp
4.0 s
2.0 s
P min
5 mbar
5 mbar
P max
35 mbar
30 mbar
Byflow
3.0 l/min
3.0 l/min
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...