9 | Alarms and troubleshooting
120
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#137
7
Expiratory
sensor?
No expiratory
flow can be meas
-
ured. The expira
-
tory flow sensor
is missing or the
measurement is
implausible.
Information:
- Volume measurement not possible
- No trigger detection possible
- No leak compensation possible
- No tube compensation possible
Please check:
- Expiratory flow sensor for proper placement
- That the expiratory flow sensor is clean and
dry
- Whether the expiratory valve needs to be
replaced
If the alarm recurs or persists:
Contact authorised service agent.
#138
1
Closed
suction
The suction
routine for closed
suction systems is
activated
To terminate the suction phase of the device,
press the <Post-oxygen.> button or the <Stop>
button.
#139
5
Open suction
The suction
routine for open
suction systems is
activated
To terminate the suction phase of the device,
reconnect the patient.
#140
4
Rate spont.
high
Rate of spontane-
ous breathing >
Alarm limit
Please check:
- Patient condition and breathing pattern
- That Auto triggering is not the cause
Consider:
- Adjusting alarm limits or ventilator settings
- Reviewing sedation protocol
#141
4–6
Rate spont.
low
Rate of spontane-
ous breathing <
Alarm limit
Please check:
- Patient condition and breathing pattern
- Suitability of the selected ventilation mode
Consider:
- Adjusting alarm limits or ventilator settings
- Reviewing sedation protocol
#142
4
Pplateau high Pplateau > Alarm
Limit
Please check:
- Patient condition and breathing pattern
Consider:
- Adjusting alarm limits or ventilator settings
#143
1
Some alarm
limits deacti-
vated
Some of the
audible alarms are
deactivated.
Information:
Ensure close monitoring!
#144
1
Ventilation
alarms
inactive
All ventilation
alarms are deac-
tivated.
Information:
- The alarm management was adapted because of
an active manoeuvre.
Ensure close monitoring!
#145
1
Risk of
trapping
Expiration time
> 2 x RCexp
Information:
- Ventilator will not adapt inspiratory time and
PS-Endflow
#146
1
Continuous
nebulization
enabled
Continuous
nebulization is
enabled.
Information:
- Switchover to continuous nebulization due to
high respiratory rate
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...