9 | Alarms and troubleshooting
146
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Code Prior-
ity
Alarm
message
Cause
Corrective measure
#512
3
CO
2
sensor
defect
The CO2 sensor
reports an error
or is missing.
Delay:
30 seconds max.
Information:
- Sensor cannot be activated
Do the following:
- Use external CO2 monitoring if required
If the alarm recurs or persists:
- Contact an authorised service agent.
#513
3
CO
2
calibra-
tion
The CO
2
sensor
needs to be
calibrated.
Delay:
30 seconds max.
Do the following:
- Rule out contamination of CO2 sensor
- Perform calibration
#514
3
CO
2
sensor
defect
The CO2 sensor
reports an error.
Delay:
30 seconds max.
Information:
- Sensor cannot be activated
Do the following:
- Disconnect CO2 sensor from serial interface,
then reconnect
- Replace defective CO2 sensor if applicable
- Use external CO2 monitoring if required
If the alarm recurs or persists:
- Contact an authorised service agent.
#515
3
Invalid CO
2
measure-
ment
The end-tidal
range or the
breath cannot be
reliably detected.
Delay:
30 seconds max.
Do the following:
- Rule out leak in breathing circuit and sampling
tube
- Watch patient to determine pathological breath-
ing patterns
- Check ventilator settings
#516
3
CO
2
sensor
adapter
No reliable read
-
ings from the
measuring cell.
Do the following:
- Check the measuring cell of the CO2 sensor or
the sampling line
- Replace defective CO2 sensor if applicable
- Use external CO2 monitoring if required
If the alarm recurs or persists:
- Contact an authorised service agent.
#517
1
CO
2
sensor
temperature
Wait for sensor
temperature to
stabilize.
Do the following:
- Wait until CO
2
sensor temperature is within
operating range.
#521
1
5
Setup error
Loading the
trend data and
setup file failed.
Information:
- Safe operation of the ventilator is ensured
- Default setup configuration is displayed
- Ventilator tries loading trend data and your
individual screen configuration automatically
- A restart may solve the problem
Do the following, if alarm recurs or persists:
- Disconnect patient from ventilator
- Continue ventilation with another ventilator
- Put this ventilator out of service
- Contact an authorised service agent.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...