A 3 | Functions
54
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
A 3.10.2
Measuring principle
The CO
2
sensor sends infra-red light of a specific wavelength through the respira
-
tory air. The absorbed portion of transmitted light is determined with a light-sen-
sitive detector and it is used for calculation of the end-tidal CO
2
concentration.
A 3.10.3
Power supply
The CO
2
sensor is powered via the serial interface.
If the main power supply fails, the sensor continues to be powered from the elisa
300 / elisa 500 unit as long as the elisa 300 / elisa 500 battery maintains the ven-
tilation function.
A 3.10.4
Breathing circuit
The CO
2
sensor is compatible with all approved breathing circuits.
A 3.10.5
Connecting a CO
2
sensor
Remark
The CO
2
sensor can be used within the patient environment.
Warning
First connect the CO
2
sensor to the device (power supply) before
you allow it to come into contact with the patient.
Warning
Be sure to connect the CO
2
sensor to one of the BF interfaces on
the back of the device.
A 3.10.6
Connecting the LeoCap CO
2
sensor
Connect the LeoCap CO
2
sensor to the configured BF interface on the back of
the ventilator using the connection cable.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...